To activate this we're gonna hold on this (On/Off Button)
And it's gonna go through its whole startup process, so these are just like I guess model numbers and stuff
Reservoir volume continuous rate, and how much is given nice.
So we do not mess with any of those settings at all? Those that had just flashed by
If they're just turning it on yeah? Yeah you
Shouldn't. No.
Now if they're setting up their pump
You would adjust the settings. So this is their their comp they just put the new cassette on.
So then they'd hit the next button
The reservoir volume, you just put a hundred in there so it needs to be 100. (A fresh cassette always has 100mLs)
Okay, so and then you hit enter
So enter to confirm next to go the next then the rate
It would probably just stay the same
They're an existing patient, or if you're titrating. If so then they would increase the rate
Depending on what the physician had asked, okay, I see
To confirm and then next to move
The amount given: you haven't given any drug so they always have to clear out the pump so they hit clear and
Then if oh no, so this is the setup so this needs to be cleared by default if we're setting and turning
So air detector is always gonna be off upstream sensor is always off
You're back to the home screen. If it's a Monday Wednesday Friday (Common cassette change schedule) then they have to prime their line
Ok so this is our the priming and then this is where we hold on the PRIME button.
You'll hear 1, 2, 3 and then let it go
Oh ok and then hold it back down again, so
You tell it to prime and then you hold on the prime button until the fluid begins to pump
And it's it's just safety mechanism, so if they roll over on it or something
That makes sense okay, so it stopped so then you have to hold it back down again
We're going to keep going until it reaches the surface of the tubing
There is cool
Take it off
So they have to start the pump before putting it on them because you want it to be a continuous
Change out so then they could start
Start and have to hold it down
Then it'll go through the settings again
And you'll be able to see what your rate is to double-check that you make sure you have to migrate
And it already that's cool. It's so it already detects how many mL is used to Prime the tubing so you know?
Yeah, it needs to it
Okay, cool, so then at this point. We'd hit yes, right so then oh, it's already going. It's going once
Once you see "RUN" at the top
It's ready to go so so the patient would have like
The old pump on them
Patient needs sort of clean hands again. Put gloves on. Yes. I tell them to get three
With gloves on
So they would step up they would take off the old pump off the line and take it off
They come in they clean that in this came the port okay, and then they take the new one
And attach it to them
And I always tell them to get three (alcohol pads) if they dropped one got you clean it
So
So the pump has to be running before they change it out
There's no break; No stoppage of medication in between changes
At this point, so when it says, "RUN" it's already running
So as you said and so this is how they'll come in when it says "RUN" and it'll say the reservoir volume
You wanted to check the settings so you wanted to know what the rate there at
Hundred mLs in there, this is the rate just 36 mLs for 24 hours and the amount given. It's all there and you can document that
from the cartridge itself. And you just hit the "NEXT" button to go through
All the settings and that's really at the end of the day without holding a button, this is the only one that's gonna be responsive
Yes. Correct. Turning it off or stopping it, you'd have to hold it down
So question if so will the pump stop right now if I were to hold on the "OFF" Button?
It has to be stopped first
Okay, and then for "LOCK" button is that like okay like I'm done, and then just lock it to make sure
The lock has nothing to do with anything so in order to stop this you'd have to hold on this and
The three lines would have to appear and that's the only way it stops
It even takes a while to do that - Wow I mean, it makes sense it's